Intinion 1.5 T MRI Scanner.
FDA Recall Z-0272-05 · Posted 2004-12-02
Advertisement
Recalling FirmPhilips Medical Systems (Cleveland) Inc
Recall StatusTerminated
Product CodeLNH
Date Initiated2003-04-17
Date Terminated2008-10-29
Product Quantity93 units.
Reason for Recall
There is a potential for the patient to be exposed to cryogenic fluids (liquid air) dripping from the vent during a quench due to the location of the vent.
Firm Action
A Mandatory Service Letter, dated April 17, 2003, was sent to service technicians instructing the technicans to perform a mandatory safety modification.
Root Cause
Other
Distribution Pattern
Throughout the United States and to Denmark, England, China, Italy, Mexico, Israel, and Puerto Rico.