Intinion 1.5 T MRI Scanner.

FDA Recall Z-0272-05 · Posted 2004-12-02

Recall StatusTerminated
Product CodeLNH
Date Initiated2003-04-17
Date Terminated2008-10-29
Product Quantity93 units.

Reason for Recall

There is a potential for the patient to be exposed to cryogenic fluids (liquid air) dripping from the vent during a quench due to the location of the vent.

Firm Action

A Mandatory Service Letter, dated April 17, 2003, was sent to service technicians instructing the technicans to perform a mandatory safety modification.

Root Cause

Other

Distribution Pattern

Throughout the United States and to Denmark, England, China, Italy, Mexico, Israel, and Puerto Rico.