Medfusion Model 4000 Syringe Infusion Pump, with software versions V1.6.0 or V1.6.1 that have implemented the Smart Pump Programming (SPP) feature in a clinical treatment environment. Used for infusion of fluids requiring precisely controlled infusion rates including blood or blood products, lipids, drugs, antibiotics, enteral solutions, and other therapeutic fluids.

FDA Recall Z-0272-2023 · Posted 2022-11-21

Recall StatusOpen, Classified
Product CodeFRN
Date Initiated2022-10-10
Product Quantity422 devices

Reason for Recall

Potential for delay in sending Infusion Start Messages to third-party Electronic Health Record systems which may affect infusion auto-programming on the pump and documentation

Firm Action

Smiths Medical notified consignees initially via telephone on 10/10/2022 and followed up with a letter via email on 10/11/2022. The letter was titled "URGENT MEDICAL DEVICE CORRECTION" and instructed consignees to locate all affected pumps and ensure all users or potential users of these devices are immediately made aware of this notification and proposed mitigations, complete and return the attached Response Form to [email protected], and if Smiths Medical does not contact you to schedule the implementation of the software updates, please contact Smiths Medical Technical Assistance. For further inquiries, please contact Smiths Medical using the following information: email: [email protected].

Root Cause

Under Investigation by firm

Distribution Pattern

US distribution to states of: Indiana, Texas and Virginia