Terminated

Precision 500D Classical R&F System with control room PC; Image-intensified fluoroscopic x-ray system; ; Model Number: 5117866-27.

FDA Recall Z-0274-2008 · Posted 2008-08-02

Recalling FirmGE Healthcare
Recall StatusTerminated
Product CodeJAA
Date Initiated2007-08-01
Date Terminated2008-09-10
Product Quantity101

Reason for Recall

Inaccurate cassette size display reading/ Inaccurate mAs reading: 1. When the device is in PBL mode and a larger film cassette is inserted into the table Bucky (replacing a smaller cassette) after a protocol is selected on the console, the collimator will not adjust to the larger cassette size. The collimator stays at the original smaller cassette size and allows the operator to take an exposure.

Firm Action

Consignees were notified by two "Product Safety Notification" letters on July 10, 2007. Each letter addressed one of the two safety issues with the affected product. The letters provided safety instructions on how avoid the software error and informed users that their representative would contact them to schedule a visit for a software correction installation.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution --USA including states of: AK, AL, AR, CA, CO, CT, FL, GA, IA, ID, IL ,IN, LA, MA, MD, MI, MO,MS, NC, ND, NE, NH, NJ, NM, NV, OH,OK, OR, PA, SC, SD, TN, TX, UT, VA, VT and WA; USA territory of: PR and countries: EGYPT, SAUDI ARABIA, MACEDONIA, FRANCE and GREAT BRITAIN.