Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Barco Surgical Display, Model MDSC-2124 The 24-inch MDSC-2124 is a near-patient surgical display that uses widescreen (16:10) LCD technology featuring full High Definition resolution (1920 x 1200). Featuring broad input connectivity, the MDSC-2124 offers a versatile display solution for HD endoscopy cameras, room and boom cameras, ultrasound, PACS and patient information.

FDA Recall Z-0276-2009 · Posted 2008-11-07

Recalling FirmBarcoview
Recall StatusTerminated
Product CodeRHF
Date Initiated2008-09-09
Date Terminated2009-06-02
Product Quantity25 units

Reason for Recall

The front protective cover may loosen and completely fall off.

Firm Action

Consignees were notified via a Field Safety Notice letter on 09/09/2008. They were instructed to inspect the units to check if the front panel is tightly secured to the display housing. Any front panel found to be lose or showing a gap between the front plate and the back shell of the monitor is to be reported the BARCO Service & Support organization and is to be secured using adhesive tape. Follow up Field Safety Notice was sent to consignees on 09/26/2008 indicating the root cause and the firm's corrective field actions implemented October 15, 2008.

Root Cause

Other

Distribution Pattern

Nationwide Distribution --- including states of AK, GA, IL, and MA.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑