Terminated

Synthes Application Instrument for Sternal ZIPFIX. Indications for use include primary or secondary closure/repair of the sternum following sternotomy or fracture of the sternum to stabilize the sternum and promote fusion.

FDA Recall Z-0276-2016 · Posted 2015-11-14

Recall StatusTerminated
Product CodeJDQ
Date Initiated2015-10-19
Date Terminated2016-08-26
Product Quantity267

Reason for Recall

Certain lots of the Application Instrument for Sternal ZIPFIX that were previously recalled on August 19, 2015 and subsequently repaired and returned to customers between August 26, 2015 and September 23, 2015. The screws on the repaired instruments have the potential to break or loosen. DePuy Synthes is requesting the return of the repaired device in exchange for a new device.

Firm Action

Communication to the accounts having record of receiving this product was sent on October 19, 2015.

Root Cause

Device Design

Distribution Pattern

Distributed in the states of KY, UT, CA, WI, OH, CO, IL, TX, OK, WA, MI, LA, MA, PA and NE.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.