Synthes External Fixation System Tube to Tube Clamp, Part Number 390.007. Intended for use to provide treatment for long bone and pelvic fractures that require external fixation. Specifically, the components can be used for: Stabilization of soft tissues and fractures; Polytrauma/multiple orthopedic trauma; Vertically stable pelvic fractures, or as treatment adjunct for vertically unstable pelvic fractures; Arthrodeses and osteotomies with soft tissue problems - failures of total joints; Neutralization of fractures stabilized with limited internal fixation; Non-unions/septic non-unions; Intra-operative reductions/stabilization tool to assist with indirect reduction; and Unilateral rectilinear bone segment transport or leg lengthening.
FDA Recall Z-0277-2013 · Posted 2012-11-13
Advertisement
Recalling FirmSynthes USA HQ, Inc.
Recall StatusTerminated
Product CodeKTT
Date Initiated2012-09-28
Date Terminated2015-09-21
Product Quantity8
Reason for Recall
The insert provided with the product for the specific lot number indicates that this product is MR safe. This product is MR Conditional and is etched as MR Conditional. The insert is incorrect.
Firm Action
Synthes sent an "URGENT NOTICE: MEDICAL DEVICE RECALL" letter dated September 28, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Customers were asked to complete and return a Verification Form via fax to 610-251-9005. Contact the firm at 610-719-5450 for questions regarding this notice.
Root Cause
Labeling mix-ups
Distribution Pattern
Nationwide Distribution-including the states of CA, MN, NC, and TN.