Medtronic Midas Rex Legend High Speed Pneumatic System

FDA Recall Z-0278-05 · Posted 2004-12-04

Recall StatusTerminated
Product CodeHBB
Date Initiated2004-11-11
Date Terminated2005-09-20
Product Quantity1432 units

Reason for Recall

Softer hose increases the propensity for outer exhaust hose to kink or become occluded as a result of clamping. Complaints of hose rupture/burst have been reported.

Firm Action

The firm initiated the recall via letter on November 11, 2004.

Root Cause

Other

Distribution Pattern

Nationwide and to the following foreign countries; Australia; Japan; Canada, China; Netherlands; Columbia; Mexico; Argentina; Sweden; Poland; Spain; Germany; Great Britan; Chile; Switzerland; Venezuela