HX2 Temp Manage SYS 120V The Terumo HX2 Temperature Management System is a selfcontained, dual channel system for supplying temperature controlled water to heat exchangers used in an extracorporeal circuit.
FDA Recall Z-0278-2013 · Posted 2012-11-13
Advertisement
Recalling FirmTerumo Cardiovascular Systems Corporation
Recall StatusTerminated
Product CodeDWC
Date Initiated2012-10-16
Date Terminated2013-07-11
Product Quantity19 units
Reason for Recall
Based on a retrospective review of quality data, AC Control boards used in the manufacture of the
HX2 Temperature Management System were identified as being manufactured using the incorrect
flux material. Use of the incorrect flux may contribute to premature board failure. A search of
TCVS complaints revealed 3 complaints received related to the HX2 failure. Each failure occurred
during set up
Firm Action
TERUMO sent an Urgent Medical Device Recall letter dated October 15, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Terumo has determined that a field correction is required to notify affected customers that the boards may fail, instruct the users that in a case of failure to swich to the second channel on the device or use a back-up unit, and that TCVS will replace the affected boards. Customers with questions were instructed to call 1-800-621-2818. For questions regarding this recall call 734-741-6173.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Nationwide Distribution including CA, HI, KS, LA, MI, MO, OR, PA, UT, and WI