Zimmer Dental GemLock RHD2.5 Hex Driver. The label shows the product is Manufactured by Zimmer Dental Inc., Carlsbad, CA. Endosseous dental implant accessories are manually powered devices intended to aid in the placement or removal of endosseous dental implants and abutments, prepare the site for placement of endosseous dental implants or abutments, aid in the fitting of endosseous dental implants or abutments, aid in the fabrication of dental prosthetics, and be used as an accessory with endosseous dental implants when tissue contact will last less than 1 hour.

FDA Recall Z-0279-2018 · Posted None

Recalling FirmZimmer Dental Inc
Recall StatusTerminated
Product CodeNDP
Date Initiated2017-10-11
Date Terminated2019-03-13
Product Quantity356 units

Reason for Recall

The device may contain a manufacturing condition affecting the geometry of the driver's hex feature which prevents it from properly engaging the fixture mount transfer and/or a dental implant.

Firm Action

The recalling firm, Zimmer Biomet, issued "URGENT MEDICAL DEVICE RECALL" letters addressed to Clinicians dated 10/11/2017 via FedEx on 10/11/2017. Letters to the distributors dated 10/11/2017 were issued via courier or email on 10/11/2017. The letters described the product, problem and actions to be taken. The customers were instructed to review notification for awareness; review your inventory; immediately return all affected product from your facility to: Field Action PM Regulatory Compliance Zimmer Biomet 4555 Riverside Dr. Palm Beach Gardens, FL 33410 US and complete and return the Certificate of Acknowledgment by fax to 574-372-4265 or email: [email protected]. If you have further questions or concerns after reviewing this notice, please call the Recall Department at 561-776-6700 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a prompt to leave a voicemail. Alternatively, your questions may be emailed to [email protected].

Root Cause

Employee error

Distribution Pattern

Worldwide Distribution: US (nationwide) and countries of: Canada, Australia, Belgium, Bulgaria, Chile, Costa Rica, Egypt, France, Georgia, Germany, Hong Kong, Hungary, Iran, Ireland, Israel, Italy, Japan, Lebanon, Lithuania, Netherlands, Nicaragua, Poland, Portugal, Romania, Serbia, Shanghai, Spain, Taiwan, Thailand, Tunisia, Turkey, UK, and United Arab Emirates.