Synapse PACS Software Version 5.6.1 - Product Usage: intended for use as a web based application on an off-the shelf PC which meets or exceeds minimum specifications and is networked with a FUJIFILM Synapse PACS server.

FDA Recall Z-0282-2021 · Posted None

Recall StatusTerminated
Product CodeLLZ
Date Initiated2020-09-11
Date Terminated2021-11-17
Product Quantity1

Reason for Recall

FUJIFILM has become aware of the possibility that certain CT studies may report a much higher 3D Sphere MAX value than expected. This issue is specific to 3D Sphere - other density tools are not affected.

Firm Action

Letter was sent to all affected customers on 09/11/2020 stating the following: FUJIFILM will offer an upgrade to Synapse PACS 7.1.100 to resolve this issue. Synapse PACS 7.1.100 is estimated to be available by the end of September. Until the patch is available, users can continue to safely use Synapse PACS by following these additional instructions: 1. Ensure that the 3D Sphere MAX value is within the spheres range. If the 3D Sphere MAX value is observed to be much higher than expected, the value should not be used when rendering a clinical decision until the issue is corrected. Customers should complete and return the Field Action Verification Form attached to the recall letter.

Root Cause

Software change control

Distribution Pattern

US Nationwide distribution including in the state of Minnesota.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.