Terminated

One Touch Ping Glucose Monitoring System Combines the functionality of OneTouch(R) Ping (TM) Insulin Pump and a OneTouch(R) Ping (TM) Meter Remote through radio frequency (RF) communication. Together, they provide an option to help make insulin delivery more discreet and flexible. Also, allows the most recent blood glucose results from the OneTouch(R) Ping (TM) Meter Remote to be automatically entered into bolus insulin calculations to cover carbohydrates in food or to correct for a high blood glucose level.

FDA Recall Z-0283-2009 · Posted 2008-11-07

Recalling FirmAnimas Corporation
Recall StatusTerminated
Product CodeLZG
Date Initiated2008-08-04
Date Terminated2008-11-12
Product Quantity251 units

Reason for Recall

display old, inaccurate values for bolus amount delivered and amount of planned bolus totals

Firm Action

On 8/6/08, the recalling firm telephoned all users to inform them of problem and the need to return the system.

Root Cause

Software design

Distribution Pattern

Nationwide Distribution

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.