Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

ReliOn Insulin syringes - 1cc, 31 G Distributed by: Can-Am Care, Corp., Alpharetta, GA

FDA Recall Z-0284-2009 · Posted 2008-11-05

Recalling FirmCovidien LP
Recall StatusTerminated
Product CodeFMI
Date Initiated2008-10-09
Date Terminated2010-07-06
Product Quantity471,000 (4710 boxes of 100)

Reason for Recall

Mislabeled: Package labeled as an insulin syringe for use with U-100 insulin contains an insulin syringe for use with U-40 insulin. Risk of overdose of insulin.

Firm Action

Tyco Healthcare Group LP (Covidien) notified Can-Am Care, Corp. by letter sent by e-mail and by Federal Express on 10/9/08. The letter requests that they immediately inventory, quarantine any remaining stock they may have of the product, return, and notify their customers. Wal-Mart conducted a consumer mailing on October 14, 2008, and is posting placards in their stores. Can-Am Care, whose name appears on the label of the product has concurrently posted a recall notice on their website.

Root Cause

Process design

Distribution Pattern

Nationwide

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑