Prism Medical C-300 Consumer Ceiling Lift; PN 330050, 330050R, 330053, 330053R, and 330057; weight specification 300 lbs. /136 kg. Lifting aid used by health care professionals to lift, position, and transfer clients or disabled family members.
FDA Recall Z-0284-2016 · Posted 2015-11-17
Advertisement
Recalling FirmErgoSafe Products, LLC (DBA) Prism Medical
Recall StatusTerminated
Product CodeFSA
Date Initiated2015-09-08
Date Terminated2017-05-04
Product Quantity561 lifts
Reason for Recall
Drive shaft and over speed arm may simultaneously fail potentially resulting in a patient falling during use
Firm Action
The firm, Prism Medical, sent a "Field Safety Notice (FSN 0008)" letter dated 9/8/2015 to all consignees by email via their sales portal. The letter describes the product, problem and actions to be taken. The consignees were instructed to call Prism Medical directly at 1-314-219-8655 and indicate that you are participating in the 300 series lift recall; provide your existing 300 series lift model number and serial number; follow the instructions provided; complete and return the Acknowledgement Form via scanned & signed to [email protected] or fax to: 314-692-7858, Attention: 300 series recall.
For further information please contact: Prism Medical Quality department at: 1-314-219-8655 or North American Director for Quality & Regulatory Affairs at 1-314-219-8614.
Root Cause
Device Design
Distribution Pattern
Worldwide Distribution: US (nationwide) and countries of: Australia, Canada, South Korea, Taiwan, Mexico, Japan, and Great Britain.