Ellex Solitarie Opthalmic Laser System, Model No: LP 4532 - Ellex Medical PTY, LTD 82 Gilbert Street, Adelaide, S.A. 5000, AUSTRALIA.

FDA Recall Z-0286-2008 · Posted 2007-12-20

Recalling FirmEllex USA
Recall StatusTerminated
Product CodeHQF
Date Initiated2007-06-13
Date Terminated2008-10-17
Product Quantity10

Reason for Recall

Electric Shock from Power Supply : The power supply has a defect which can cause an electrical shock to user.

Firm Action

Ellex, Inc. called each customer on 6/13/07, about the problem (defect which can cause electric shock) and advised that they would be performing modification of the power supply in the laser system and scheduling a field service visit to complete the work. Upon completion of the modification, the service technician completed the service record and left a copy with each client and informed them of the completed work. A copy of each completed service record was sent to Ellex Medical Pty. Ltd. On 9/27/07, Ellex sent a letter to all consignees, informing them that all modifications by service technicians' have been completed, and that their Solitaire LP4532 laser and/or Intregre Duo LP1RG is up to the manufacturers specifications and is in good operating order.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution - USA including states of WI, PA, CO, CA, NC, NY, SC, MD, and MS, and countries of Canada and Brazil.