Terminated

Special Orthopaedic Solution Peri-Acetabular Prosthesis V40 Femoral Extension Component Product Usage: The PAR extension piece is a component of the PAR assembly. The Peri-Acetabular Reconstruction (PAR) endoprosthesis was developed to meet the unique needs of patients who require reconstruction of large acetabular defects after a Type I and II pelvic resection.

FDA Recall Z-0286-2013 · Posted 2012-11-13

Recall StatusTerminated
Product CodeJDI
Date Initiated2012-07-16
Date Terminated2015-09-09
Product Quantity301 units

Reason for Recall

Stryker has become aware that insufficient taper torsional strength may result when a PAR extension piece is used with a forged CoCr stem.

Firm Action

Stryker Orthopaedics sent an Urgent Product Recall letter dated July 16, 2012 via Fed Ex with return receipt to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter requested customers to contact their Stryker Sales Representative to arrange for return of the affected product. Customers were instructed to fax back the attached Product Recall Acknowledgement Form to 201-821-6069. For questions call 201-972-2100.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution - US Nationwide

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑