Hurricane RX Biliary Balloon Dilatation Catheter Catalog REF No. M00545890; M00545900; M00545910; M00545920; M00545930; M00545940; M00545950; M00545960 The device is a double lumen catheter with a balloon mounted at the distal tip. Dilatation balloon catheters are used to exert radial force to dilate narrow duct segments, as well as the Sphincter of Oddi.
FDA Recall Z-0286-2021 · Posted None
Advertisement
Recalling FirmBoston Scientific Corporation
Recall StatusOpen, Classified
Product CodeFGE
Date Initiated2020-09-16
Product Quantity15,062 devices distributed in US. 12,930 distributed OUS
Reason for Recall
RX tunnel component (black sheath) may detach from the catheter shaft due to lower bond strength.
Firm Action
All consignees were sent a copy of the customer notification letter via Federal Express Priority Overnight mail on 09/16/2020. The notification instructs the customer to check their inventory, remove any affected product from their inventory, and return it to BSC. The notification also requests the completion of a Reply Verification Tracking Form as a method of documenting the presence or absence of affected product from their inventory.
Follow-up communications will be made to the accounts that do not respond to this inquiry.
The returned products will be quarantined until an appropriate course of action has been determined. BSC will notify FDA of this decision prior to its implementation.
Root Cause
Vendor change control
Distribution Pattern
Worldwide distribution - US Nationwide and International distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.