NAMIC Custom Angiographic Kit, Right Heart Kit, REF/Catalog # 60190254, Sterile, and bulk Non-Sterile Manifolds Assemblies, Product #s 40338200 and 40336200 (foreign product for further processing, e.g., packaging, labeling, sterilization, and distribution under BSC). Manufactured by: Boston Scientific Corporation/NAMIC Technology Center/Glens Falls, New York 12801. Angiographic Manifolds are intended for use in fluid management and/or invasive pressure monitoring systems. Procedures in which angiographic manifold may be used include hemodynamic pressure monitoring and intra-arterial and intravenous administration of water based solutions or radiographic contrast media.
FDA Recall Z-0287-2009 · Posted 2008-11-09
Advertisement
Recalling FirmNavilyst Medical
Recall StatusTerminated
Product CodeDTZ
Date Initiated2008-08-06
Date Terminated2009-06-16
Product QuantityDomestic - 20 units; Foreign - 16,200 bulk units
Reason for Recall
Firm became aware that a specific batch of the product may contain loose plastic particulates in the fluid pathway.
Firm Action
On 8/12/08, Navilyst Medical sent a fax and phoned the sole domestic hospital account that had received the affected units. Navilyst Medical instructed the consignee to immediately locate, segregate, and quarantine the units. The hospital consignee confirmed that all units were still in their control and that they will maintain them in quarantine awaiting return to Navilyst Medical. The sole international account, Boston Scientific in Ireland was notified of the problem on 8/06/08. This was followed up with an Urgent Medical Device Recall Letter and Verification Form sent via FedEx on 8/20/08.
Root Cause
Process design
Distribution Pattern
Worldwide Distribution --- USA including state of MO and country of Ireland.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.