Biomet 2.0 MM Stainless Steel (SS) Crimp Sleeve, sterile, use w/2.0 MM S.S. cable # 350800, Biomet Orthopedics, Inc., Warsaw, IN; REF. 350805. Intended for general orthopedic repairs including long bone fractures, bone grafting, reinforcement of bone and reattachment of the greater trochanter.

FDA Recall Z-0288-2009 · Posted 2008-11-09

Recalling FirmBiomet, Inc.
Recall StatusTerminated
Product CodeJDQ
Date Initiated2008-09-24
Date Terminated2009-06-29
Product Quantity48

Reason for Recall

The outer package is properly labeled, but the inner package may be labeled as a femoral component.

Firm Action

The firm's sales force was notified of the problem via an Urgent Medical Device Removal Notice letter dated 9/24/08, and provided instructions to retrieve the product from customer accounts and to provide them with a copy of the removal letter. Distributors were to immediately locate and remove the identified implants from circulation, carefully follow the instructions, and fax a copy of the Response Form prior to the return of the product. If the product was further distributed, hospital personnel MUST be notified via the enclosed "Dear Biomet Customer" notice. The letter MUST be given to hospital personnel responsible for receiving recall notices but distributors are charged with the location and return of those products.

Root Cause

Labeling mix-ups

Distribution Pattern

Worldwide Distribution --- USA including states of Arkansas, Florida, Kansas, New Jersey, and Ohio, and countries of Argentina, Canada, Chile Italy, Mexico and United Kingdom.