Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Instructions For Use document of the GPS Trackers.

FDA Recall Z-0288-2021 · Posted None

Recalling FirmBlue Ortho
Recall StatusTerminated
Product CodeOLO
Date Initiated2020-09-10
Date Terminated2023-05-31
Product Quantity3471 units

Reason for Recall

Appropriateness of the language used in the Tracker reprocessing instruction. Reason is that some Low Temperature sterilization conditions mentioned in the IFUs are approved based on rational documents.

Firm Action

On 9/10/2020 the firm sent an Urgent Field Safety Notification to its consignees US FDA expressed concern with the appropriateness of the language used in the Tracker reprocessing instruction. Reason is that some Low Temperature sterilization conditions mentioned in the IFUs are approved based on rational documents, leveraged from proper validation files. Based on this concern, Blue Ortho decided to revise the appropriateness of the language used in the Tracker reprocessing instruction LEFUS Tracker for Low Temperature Sterilization v1.0 to remove such conditions. Blue Ortho has already updated the GPS Trackers IFUs, to remove the sterilizations cycle based on rational and not effectively supported by validation. The new version of the IFU is LEFUS Tracker for Low Temperature Sterilization v1.2 (version 1.1 has not been released to the field). The LEFUS Tracker for Low Temperature Sterilization v1.2 is enclosed to this notice. Transmission of this Field Safety Notice: " This notice needs to be passed on all those who need to be aware within your organization or to any organization where the potentially affected devices or documents have been transferred. " Please transfer this notice to other organizations on which this action has an impact. " Please maintain awareness on this notice.

Root Cause

Other

Distribution Pattern

Domestic Distribution Only: CA, CO, SC and WI.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑