ImmuneTech brand MyAllergyTest, Specimen Collection Kit, Part No. 8015, Screening Pack: 50 test (with lancet), Product is manufactured distributed by ImmuneTech, Inc., Foster City, CA; but the non-sterile alcohol prep pads are manufactured by H & P Industries, Inc., Triad Group, Hartland, WI The MyAIIergyTest is for the collection and transport of a capillary blood sample to a clinical laboratory for allergen specific lgE profile testing.
FDA Recall Z-0289-2012 · Posted 2011-11-25
Advertisement
Recalling FirmImmunetech Inc
Recall StatusTerminated
Product CodeJKA
Date Initiated2011-11-04
Date Terminated2012-06-14
Product Quantity50
Reason for Recall
ImmuneTech is recalling certain Specimen Collection Kits that contain Triad brand non-sterile alcohol prep pads which have been recalled by Triad Group, Hartland, WI due to Bacillus cereus, microbial contamination.
Firm Action
The firm, IMMUNETECH, sent an "URGENT PRODUCT RECALL NOTIFICATION" letter dated November 4, 2011, to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to inspect their inventory and quarantine all affected kits; if they have further distributed the affected kits, contact Immunetech to accurately identify and immediately notify their customers and facilitate the return and exchange of all remaining kits.
If you have any questions, call (650) 312-1066, Monday-Friday, 9:00am -5:00pm (pacific) or email: [email protected].
Root Cause
Material/Component Contamination
Distribution Pattern
Worldwide distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.