Terminated

Candela Smoothbeam Laser System Model Numbers: 9914-00-0820 9914-06-0820 9914-07-0820 9914-08-0820 9914-10-0820 Units are the same except for color

FDA Recall Z-0291-04 · Posted 2004-01-01

Recall StatusTerminated
Product CodeGEX
Date Initiated2003-11-25
Date Terminated2006-12-06
Product Quantity506 units

Reason for Recall

Laser may pulse and not dispense cryogen resulting in patient injury

Firm Action

Candela notified direct customers and distributors by letter dated 11125/03 and issued Certified Mail. Accounts are requested to perform two tests on the unit to determine if the unit has the problem. If a unit fails, Candela's Field Service will make the repair. Distributors are requested to notify their customers.

Root Cause

Other

Distribution Pattern

Nationwide Foreign: Canada,Thailand, Germany, Spain , Japan, Italy, Greece, Saudi Arabia, UK, France, Switzerland, Mexico, Ukraine, Kuwait, Netherlands.