Pioneer Quantum Spinal Rod System, MIS Yoke Manipulator, Long Slot, non-sterile; Cat # 51-YOKEMAN-L, Pioneer Surgical Technology, Marquette, MI 49855

FDA Recall Z-0291-2008 · Posted 2007-11-27

Recall StatusTerminated
Product CodeHWR
Date Initiated2007-10-10
Date Terminated2008-03-31
Product Quantity38

Reason for Recall

The pins may come out of the instrument and, if this happens, they could fall into/remain in the patient.

Firm Action

Consignees were notified via a Product Recall letter dated 10/10/07 to remove these instruments from their instrument cases and to return them to the firm.

Root Cause

Component change control

Distribution Pattern

Nationwide: including states of California, Florida, Michigan and North Carolina.