Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Philips IntelliVue MX40 Patient Monitor: IntelliVue MX40 Smart-hopping (2.4GHz) Product Number: 865351 Worldwide Except USA Exchange part: 453564262491 453564262511 453564615311 453564615331 453564262571 453564262591 Exchange part: 453564262531 453564262551 453564262571 453564262591 453564615351 453564615371 453564467761 (Korea) 453564467781 (Korea) 453564451791 (Japan, Taiwan, Indonesia) 453564451811 (Japan, Taiwan, Indonesia) 453564467821 (Mexico) 453564467841 (China) 453564467861 (China) Worldwide except for the USA SW Revisions: B.05.28, B.05.29, and B.05.32 Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals

FDA Recall Z-0292-2017 · Posted None

Recall StatusTerminated
Product CodeMHX
Date Initiated2016-09-19
Date Terminated2018-05-25
Product Quantity1824 units

Reason for Recall

Philips IntelliVue MX40 Patient Wearable Monitor Configuration Setting Disables Generation and Delivery of ECG Alarms to PIIC iX

Firm Action

The firm, Philips Healthcare, sent a "Field Safety Notice (FSN)" dated 9/14/2016 to its customers. The notice described the product, problem and actions to be taken. The customers were instructed to make sure that the MX40 Patient Wearable Monitor local configuration setting 'Alarms On' is configured to its factory default until your software is upgraded; and follow the instructions listed in the letter. A Philips Healthcare representative will contact customers with affected devices to arrange for the installation of the software upgrade. If you need any further information or support concerning this issue, please contact Philips Customer Support at 1-800-722-9377.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution-US (Nationwide) and countries of: Aruba, Australia, Austria, Belgium, Bulgaria, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Ghana, Greece, Hong Kong, Iceland, India, Iraq, Ireland, Italy, Japan, Korea, Republic of, Malaysia, Mexico, Netherlands, New Zealand, Norway, Philippines, Poland, Portugal, Singapore, Slovenia, South Africa, Spain Sweden, Switzerland, Thailand, Turkey, United Arab Emirates and United Kingdom.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑