Ingenia 1.5T and 3.0T MRI Systems, Model Numbers: 781396 and 781377, 510(k) #K110151. System units equipped with software version: 4.1.1 are subject to recall/field correction. This MRI system is indicated for use as a diagnostic device. It can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body, or extremities. Image appearance is determined by the spatial distribution and flow as well as by many different NMR related properties of the tissue and anatomy studied and the MRI scan technique applied. These images and/or spectra, when interpreted by a trained physician, provide information that may assist the diagnosis.

FDA Recall Z-0293-2012 · Posted 2011-11-28

Recall StatusTerminated
Product CodeLNH
Date Initiated2011-08-03
Date Terminated2013-02-22
Product Quantity26 units - 12 units distributed in U.S. & 16 distributed to foreign customers

Reason for Recall

The firm determined that during scanning procedures, for some larger patients, some reddening of the skin of can occur in certain instances due to RF heating when the torso of the patient is positioned within the body coil.

Firm Action

Philips Medical Systems ( Cleveland) Inc. sent a Urgent Field Safety Notice dated August 3, 2011, to all affected customers. Philips Field Service Engineers were dispatched to the affected customer sites to deliver the recall notification and to perform the necessary software corrections. The letter lists the affected MRI systems; describes the problem; the actions to be taken by the customer/user; it also informs the customer of the upcoming software correction which will be applied to each unit by the Philips Field Service Engineers in order to address this issue. If any additional information is required, the customer is instructed to contact their local Philips Field Service Engineer. For further questions please call Customer Care Solutions Center (CCSC) at 800-722-9377.

Root Cause

Software design