LIFEPAK 20 defibrillator/monitor. Paddle sets, an optional accessory item, used with the LP 20.

FDA Recall Z-0294-04 · Posted 2004-01-06

Recall StatusTerminated
Product CodeMKI
Date Initiated2003-11-12
Date Terminated2004-02-25
Product Quantity180 sets

Reason for Recall

Device unable to recognize the connection to the hard paddles accessory.

Firm Action

Beginning on 11/12/03 representatives from the firm visited the consignees and replaced the paddles. A letter dated November 2003 was provided to the consignees.

Root Cause

Other

Distribution Pattern

The firm distributed devices to 32 hospitals and medical centers located throughout the United States.