Terumo Advanced Perfusion System 1 Electronic Oxygen Blender/analyzer; Catalog number 801188.

FDA Recall Z-0294-2007 · Posted 2006-12-19

Recall StatusTerminated
Product CodeDTQ
Date Initiated2004-01-01
Date Terminated2007-02-02
Product Quantity199 units

Reason for Recall

The flow meter could malfunction resulting in loss of touchscreen control of the gas system, but the alternate mechanical control will remain operable.

Firm Action

U.S. consignees were visited by firm representatives between January 2004 and September 2005 and the device modified to correct the problem. Approximately 32 exported units are still in need of being upgraded.

Root Cause

Other

Distribution Pattern

Worldwide, including USA, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.