Medtronic 0-arm Imaging System, Product Catalog Number: B1-700-00027, Medtronic Navigation, Inc., Littleton, MA 01460 (software version 3.0)
FDA Recall Z-0294-2008 · Posted 2007-11-27
Advertisement
Recalling FirmMedtronic Navigation, Inc
Recall StatusTerminated
Product CodeJZL
Date Initiated2007-10-17
Date Terminated2007-11-27
Product Quantity7 units
Reason for Recall
Reversed Instrument Position : a defect in the software version 3.0 causes an error in the position of surgical instruments and relative motion are reversed 180' on the StealthStation image. (If the StealthStation is not used, or if the patient orientation is prone or supine, the problem does not occur)
Firm Action
Medtronic Navigation notified customers by letter on October 17, 2007. Medtronic Field service will perform upgrade (software version 3.0.1) at user site.
Root Cause
Software change control
Distribution Pattern
Worldwide: USA including states of CO, IN, NE, and MN and countries of Canada and Germany