Terminated

Fusion Workstation.; Indicated for the transmission and review of radiological images.

FDA Recall Z-0294-2017 · Posted None

Recall StatusTerminated
Product CodeLLZ
Date Initiated2011-10-28
Date Terminated2016-12-14
Product Quantity13 sites potentially hae the affected version

Reason for Recall

After a period of time running Fusion Workstation, the Hounsfield measurement tool will report incorrect values.

Firm Action

The recalling firm initiated phone calls to affected sites beginning 1/30/2016.

Root Cause

Software design

Distribution Pattern

Distribution was made to AZ, CA, ID, IL, IN, MI, MN, and NJ.