RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6 and RayStation 7, a stand-alone software treatment planning system.
FDA Recall Z-0294-2019 · Posted None
Advertisement
Recalling FirmRAYSEARCH LABORATORIES AB
Recall StatusTerminated
Product CodeMUJ
Date Initiated2018-09-21
Date Terminated2020-06-16
Product Quantity110 units
Reason for Recall
Robust optimization for plans using beam set + background dose is not supported in RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6 and RayStation 7.
Firm Action
Ray Search Labs notified userson 9/21/18 by email advising users of the action to be taken: After creating an adaptive plan from a plan with robust objectives, remove all robust objectives that are automatically added to the adaptive plan and replace them with other appropriate objectives before optimizing the adapted plan.
o Do not use an optimization function template with robust functions for a dependent beam set.o Do not use scripting to add robust optimization functions related to beam set + background dose. Reply form to be completed/signed and returned.
Regulatory information, please contact [email protected]
Root Cause
Software design
Distribution Pattern
Nationwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source