POLYMEDCO ULTRA-CRP High Sensitive Assay For the DIMENSION(R) systems, Catalog No. USC300. Contents: 4 x 50 Tests. For In Vitro Diagnostic (IVD) use only. COMMON/USUAL NAME: High-Sensitive C-Reactive Protein --- CLASSIFICATION NAME: C-reactive Protein, Antigen, Antiserum.
FDA Recall Z-0295-04 · Posted 2004-01-06
Advertisement
Recalling FirmPolymedco, Inc
Recall StatusTerminated
Product CodeDCK
Date Initiated2003-12-03
Date Terminated2005-04-20
Product Quantity165 Kits
Reason for Recall
Firm was notified by the foreign manufacturer that certain lots of Ultra-CRP are prone to prozone (a condition of antibody antigen excess) and do not meet the labeled specifications.
Firm Action
By phone on 12/3/03 and by mail on 12/5/03.
Root Cause
Other
Distribution Pattern
Medical facilities, physicians, diagnostic laboratories, and distributors located nationwide.