Fresenius 132 Gallon Dry Acid Dissolution Unit, P/N 160147. For Use in the Preparation of acid concentrate for hemodialysis.
FDA Recall Z-0295-2016 · Posted None
Advertisement
Recalling FirmFresenius Medical Care Renal Therapies Group, LLC
Recall StatusTerminated
Product CodeKPO
Date Initiated2015-10-23
Date Terminated2021-03-02
Product Quantity348 units
Reason for Recall
Dissolution units and replacement motors units do not provide full thermal protection for both the Main and Starter coils resulting in the generation of smoke and/or potentially lead to fire.
Firm Action
Fresenius issued an Urgent - Medical Device Correction notification, dated October 21, 2015, on 10/27/15. The letter identified the affected device and the reason for the recall. It also contained instructions (on how to perform the required wiring modification) along with a Reply Form. Customers are to return the form after completing the modification. Inquiries regarding the correction should be directed to Fresenius Technical Services at 800-227-2572, referencing the Field Action number FA-2015-01-W.
Root Cause
Device Design
Distribution Pattern
Nationwide Distribution.