Xoran MiniCAT for ENT, Xoran Technologies, Incorporated, Ann Arbor, MI 48103. This device is a cone beam CT scan for sinue, temporal bones and maxillofacial imaging.

FDA Recall Z-0296-2008 · Posted 2007-11-27

Recall StatusTerminated
Product CodeJAK
Date Initiated2006-07-13
Date Terminated2008-03-13
Product Quantity5

Reason for Recall

The hard drive may crash, resulting in loss of data from patient CT scans.

Firm Action

Consignee accounts were visited between 7/25/06 and 8/10/07 and the hard drives on their systems were replaced.

Root Cause

Finished device change control

Distribution Pattern

Nationwide Distribution: California, South Dakota, Tennessee, Utah and West Virginia.