Bard electrophysiology SSV H. Lee 9.5F split sheath with valve and sideport. The product is packaged in individual pouches and shipped 5 pouches per case.

FDA Recall Z-0297-06 · Posted 2005-12-17

Recall StatusTerminated
Product CodeDYB
Date Initiated2005-11-07
Date Terminated2006-06-14
Product Quantity445 kits

Reason for Recall

mislabeled as 9.5F actually contains 10.5F

Firm Action

The recalling firm issued letters to their direct accounts via fax on 11/7/05. Letters to the hospital accounts were mailed via UPS overnight on 11/14/05.

Root Cause

Other

Distribution Pattern

The products were shipped to one distributor in CA and the label holder in MA. The products were further distributed to hospitals in CA, CO, IL, KY, MO, MS, NC, NE, NY, OH, PA, SC, and TX.