St. Jude Medical Merlin PCS Programmer, used in combination with St. Jude Medical bradycardia and tachycardia devices. Model 3650

FDA Recall Z-0297-2007 · Posted 2006-12-19

Recalling FirmSt Jude Medical CRMD
Recall StatusTerminated
Product CodeKRG
Date Initiated2006-10-06
Date Terminated2012-03-26
Product Quantity6,290

Reason for Recall

St. Jude Medical has identified a low-frequency anomaly in the software used in the APS III Model 3500/3510 and Merlin PCS Model 3650 programmers that can lead to incorrect reporting of battery voltage, expected battery longevity and Elective Replacement Indicator (ERI) status.

Firm Action

The notification material was sent to physicians by FedEx. The advisory notification was sent by FedEx on Oct. 12, 2006, to arrive at the physicians' offices on Oct. 13th. In the US, acknowledgement of receipt of the notification material will be monitored by St. Jude Medical CRMD in Sylmar, California. For the rest of the world it will be by various St. Jude offices and local distributors with this information being fed back to St. Jude Medical CRMD in Sylmar. In addition, St. Jude Medical issued a press release on October 13, 2006. Plus information concerning this advisory can be found on St. Jude's website, www.sim.com Completion of this advisory is estimated to be three months after its initiation.

Root Cause

Other

Distribution Pattern

Worldwide in USA, ARGENTINA, AUSTRALIA, AUSTRIA, BARBADOS, BELGIUM, BERMUDA, BOLIVIA, BRAZIL, CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, CYPRUS, CZECH REPUBLIC, DENMARK, ESTONIA, EUROPE, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, GUATEMALA, HONG KONG, ICELAND, INDIA, ISRAEL, ITALY, JAPAN, JORDAN, SOUTH KOREA, LATVIA, LICHTENSTEIN, LUXEMBURG, MALAYSIA, MEXICO, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, NORWAY, PORTUGAL, SAUDI ARABIA, SLOVAKIA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TUNISIA, UNITED KINGDOM & VENEZUELA