Terminated

Siemens Artis one systems with a Display Port transceiver-Fluoroscopic X-Ray System Model: 10848600 - Product Usage: he examination table as an integrated part of the system can be used for X-ray imaging, surgery and interventions. The Artis one can also support the acquisition of position triggered imaging for spatial data synthesis.

FDA Recall Z-0298-2021 · Posted None

Recall StatusTerminated
Product CodeOWB
Date Initiated2020-09-21
Date Terminated2021-08-03
Product Quantity22 units

Reason for Recall

Potential malfunction in which the examination room monitor may become intermittently dysfunctional (e.g., no display/flicking/wrong resolution).

Firm Action

Siemens issued Urgent Medical Device Correction notification letter dated 9/22/20 via AX053/20/S stating reason for recall, health risk and action to take: Siemens will correct the error with a hardware replacement of the Display Port transceiver via Update Instruction AX052/20/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the correction, the cause is eliminated, and any recurrence of this potential fault is prevented. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document.

Root Cause

Component design/selection

Distribution Pattern

US Nationwide distribution including in the states of AL, AR, AZ, CA, FL, IN, LA, MI, MO, NC, NJ, NY, PA, TX, WA.

Official Source

View FDA Recall →