Terminated

Ventana Medical System's Symphony Staining System, KPA, Model #: 2200000, Catalog #: 900-SYM1 & 900-SYM3, Ventana Medical Systems, Tucson, AZ

FDA Recall Z-0300-2008 · Posted 2008-01-09

Recall StatusTerminated
Product CodeKEY
Date Initiated2007-09-06
Date Terminated2008-07-08
Product Quantity45 units

Reason for Recall

Shock hazard: Devices released prior to full testing; the configuration of the Symphony instrument during Hi pot testing procedures did not completely evaluate all components in the test loop. In specific circumstances, a shock hazard exists.

Firm Action

On 9/6/07, the firm began visiting customers to perform the HiPot test. Prior to performing the High Potential Test of Symphony Instruments already in the Field, the Field Service Engineer or FSE ensures that the customer fully understands the purpose of the test and why it is necessary to perform the test at their site. The FSE conveys the following points regarding the HiPot test: - The HiPot test is a safety verification for the electrical path within the instrument. - The HiPot test was performed incorrectly during the manufacturing process. - The HiPot test is used to verify that if a high voltage overload condition were to occur it would not lead to a short/arc within the instrument, which could pose a risk to anyone that touches the instrument - The HiPot test places a voltage of 2,000VDC onto the high voltage path of the instrument and if any of the insulating components are compromised, the tester will notify the FSE of the condition. - If the instrument fails the HiPot test, the faulty component will be replaced and the instrument will be retested until it passes. - Due to the high voltage used the during the test, no one will be allowed to touch the instrument until the test is complete. The FSE performing the test documents that the customer was informed of the reason for the High Potential Test on their Field Service Report.

Root Cause

Process control

Distribution Pattern

Worldwide: USA, Japan and France