Terminated

Streamline TL Navigation System, Catalog Nos. 02-C-TAP-55-MDN 02-C-TAP-60-L-MDN 02-C-TAP-65-L-MDN 02-C-TAP-65-MDN 02-C-TAP-75-L-MDN 02-C-TAP-85-L-MDN 02-C-TAP-85-MDN 02-C-TAP-95-L-MDN 02-RS-INS-C-MDN 02-RS-INSERTER-MDN 02-SCREWINS-CL-MDN 02-SCREWINS-C-MDN 02-S-FINDER-L-MDN 02-SFINDER-MDN 02-S-FINDER-T-MDN 02-TAP-45-MDN 02-TAP-55-MDN 02-TAP-60-L-MDN 02-TAP-65-L-MDN 02-TAP-65-MDN 02-TAP-75-L-MDN 02-TAP-75-MDN 02-TAP-85-L-MDN 02-TAP-85-MDN 02-TAP-95-L-MDN 02-MODHANDLE-MDN Product Usage: intended to be used during the preparation and placement of screws from the Streamline TL and MIS Systems and preparation of screws from the Streamline OCT System during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures.

FDA Recall Z-0300-2021 · Posted 2020-10-27

Recall StatusTerminated
Product CodeOLO
Date Initiated2020-09-24
Date Terminated2024-09-16
Product Quantity420

Reason for Recall

The Streamline Navigation System taps have experienced binding with other devices.

Firm Action

On September 24, 2020, the firm distributed the Urgent Medical Devices Recall letter to customers. Customers were instructed to remove affected product from all user sites and inventory locations. These instrument systems are to be returned to RTI Surgical along with a completed copy of the response form. Accounts who have further distributed the affected products are also asked to immediately notify their customers by providing them a copy of the recall letter and the response form for completion when returning the affected devices to RTI Surgical.

Root Cause

Device Design

Distribution Pattern

US Nationwide distribution including in the state of Michigan.