Roche Diagnostics URISYS 1100 Urine Analyzer Part Number: 03617556001 Product Usage: The URISYS 1100 Urine Analyzer is a reflectance photometer designed to automatically read and evaluate the results of Chemstrip 5 OB, Chemstrip 7, Chemstrip 10 MD, and Chemstrip 10 UA urine test strips for various urine analytes.: specific gravity, pH, leukocytes, nitrite, protein, glucose, ketones, urobilinogen, bilirubin and blood
FDA Recall Z-0301-2019 · Posted None
Advertisement
Recalling FirmRoche Diagnostics Corporation
Recall StatusTerminated
Product CodeKHE
Date Initiated2018-09-24
Date Terminated2022-02-09
Product Quantity14,398 units
Reason for Recall
Limits of Detection (LoD) for protein, nitrite, leukocytes, and erythrocytes on the Urisys 1100 urine analyzer with Chemstrip 5 OB, Chemstrip 7, Chemstrip 10 MD, and Chemstrip 10 UA test strips were higher than what is listed in the test strip method sheets may lead to false negative results for the affected parameters
Firm Action
Roche Diagnostics notified Consignees by the Urgent Medical Device Correction (UMDC) sent UPS Ground (receipt required) on 9/24/18. For consignees associated only with a PO number address, the UMDC will be sent by USPS mail.
Consignee Instructions:
1. When using Chemstrip 5 Ob and/or Chemstrip 7 test strips, results should
only be read visually.
These test strips can no longer be read on the Urisys 1100 urine analyzer.
Follow the directions in the Instructions for Manually Reading Chemstrip test Strip results section of the UMDC.
2. When using Chemistrip 10 MD or Chemstrip 10 UA test strips:
You can choose to read the test strips visually or
If you continue to use the strips with the Urisys 1100 urine analyzer, a
negative result for any of the four affected parameters (protein, nitrite,
leukocytes, and erythrocytes) must be repeated with a new test strip
that is read visually. Follow the directions in the instructions for
Manually Reading Chemstrip Test Strip Results section of the UMDC.
3. If you are unable to implement the visual read-only method for
Chemstrip 5 OB, Chemstrip 7, Chemstrip 10 MD, or Chemstrip 10 UA test
strips:
Dispose of the affected test strip vials per local guidelines.
Contact your distributor partner for credit.
4. Complete all sections of the enclosed fax back forms
Contact Roche Diagnostics Point-of-Care Technical Service, 24 hours a day, seven days a week at 1-800-428-4674 if you have questions
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.