Infinity Docking Station (IDS) (transportable patient monitor) Repair Kit, part number 7262814
FDA Recall Z-0303-2007 · Posted 2006-12-19
Advertisement
Recalling FirmDraeger Medical, Inc.
Recall StatusTerminated
Product CodeMHX
Date Initiated2006-11-17
Date Terminated2008-11-19
Product Quantity23,345 units - 4336 US
Reason for Recall
Top cover of assembly may separate. The situation can result in the monitor falling, which could result in an injury to the clinician or patient.
Firm Action
The recaling firm issued a recall letter to the US customers 11/17/2006 via certified mail. An authorized service representative will inspect each potentially affected unit. If a unit is found to have an affected docking station, the unit will be repaired or replaced free or charge. Notification to international customers will occur through the international subsidiaries.
Root Cause
Other
Distribution Pattern
Nationwide
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.