Histochemical PAS (Periodic Acid Schiff) Reaction Set. in vitro diagnostic. Product size configuration: Set of 4 bottles including Schiff Reagent 225mL, Light Green SF Yellowish Stain 225mL, Sodium Carbonate Solution 225mL, and Periodic Acid Solution 225mL. Each kit provides reagents sufficient for a maximum of 100 tests.

FDA Recall Z-0305-2007 · Posted 2006-12-27

Recalling FirmEMD Chemicals Inc
Recall StatusTerminated
Product CodeKKS
Date Initiated2006-07-25
Date Terminated2008-10-06
Product Quantity74 kits

Reason for Recall

Due to loose caps, the SO2 levels dropped significantly from release levels causing the product not to perform properly.

Firm Action

On July 27, 2006 EMD contacted its direct account customers via letter. The firm requested discontinue use of the product and contact EMD for return.

Root Cause

Other

Distribution Pattern

Nationwide.