Performix X-ray Tube - Performix ADV, Performix Ultra, Solarix 630.

FDA Recall Z-0306-04 · Posted 2004-01-08

Recall StatusTerminated
Product CodeIYB
Date Initiated2003-12-08
Date Terminated2005-11-15
Product Quantity400

Reason for Recall

Incorrect mounting hardware may cause X-ray tube to separate from the gantry and fall.

Firm Action

Field Modification Instructions (FMI''s) will be issued to the field to inspect and locate suspect tubes and replace any that may be susceptible to failure.

Root Cause

Other

Distribution Pattern

Distributed nationwide throughout the US and worldwide.