Performix X-ray Tube - Performix ADV, Performix Ultra, Solarix 630.
FDA Recall Z-0306-04 · Posted 2004-01-08
Advertisement
Recalling FirmGeneral Electric Med Systems
Recall StatusTerminated
Product CodeIYB
Date Initiated2003-12-08
Date Terminated2005-11-15
Product Quantity400
Reason for Recall
Incorrect mounting hardware may cause X-ray tube to separate from the gantry and fall.
Firm Action
Field Modification Instructions (FMI''s) will be issued to the field to inspect and locate suspect tubes and replace any that may be susceptible to failure.
Root Cause
Other
Distribution Pattern
Distributed nationwide throughout the US and worldwide.