Terminated

CELL-DYN Sapphire Hematology Analyzer, List numbers 08H00-01, 08H00-03 A multi- parameter, automated hematology analyzer designed for in vitro diagnostic use in counting and characterizing blood cells.

FDA Recall Z-0306-2009 · Posted 2008-11-21

Recalling FirmAbbott Laboratories
Recall StatusTerminated
Product CodeGKZ
Date Initiated2008-08-25
Date Terminated2011-01-06
Product Quantity958 units

Reason for Recall

Exposed wire resulted in minor electrical shock and burn to the service technician

Firm Action

Recall issued August 22, 2008. Technical service bulletin was issued for all field personnel via internal database. No customer communication is planned as customers are instructed *not* to remove the cover, which is necessary to access the area.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution --- including USA and countries of Argentina, Australia, Brazil, Canada, Chile, China WFOTC, Colombia, Germany, Hong Kong, Japan, Mexico, New Zealand, Panama, Paraguay, Peru, Singapore, South Korea, Taiwan and Venezuela.