Elekta Leksell Gamma Plan Product Usage: Gamma Plan is a computer based system designed for Gamma Knife treatment planning.
FDA Recall Z-0307-2012 · Posted 2011-11-30
Advertisement
Recalling FirmElekta, Inc.
Recall StatusTerminated
Product CodeIWB
Date Initiated2011-06-23
Date Terminated2012-01-31
Product Quantity2
Reason for Recall
There is a risk that the user may interpret the instructions in the manual too literally, and adjust the TMR 10 dose description to obtain an ETCPD that matches previous clinical prescriptions, which in turn could lead to a prescription dose that deviates a few percent from the intended dose.
Firm Action
Elekta sent an IMPORTANT NOTICE letter dated June 23, 2011 to affected customers. The notice identified the product, problem, and corrective actions to be taken by the user. The letter advise all customers to follow any instructions or recommendations covered in the Notice. The letter states that Elekta is working on a new software version and customers will receive the future version when released. For questions contact your local Elekta representative.
Root Cause
Labeling False and Misleading
Distribution Pattern
Nationwide Distribution (USA) - including the states of: NM and WA