HeartStart FR2+ Defibrillator

FDA Recall Z-0308-2007 · Posted 2006-12-27

Recall StatusTerminated
Product CodeMKJ
Date Initiated2006-12-19
Date Terminated2011-06-22
Product Quantity2000 units

Reason for Recall

Potential for shock treatment to be delayed or prevented due to board contamination that causes intermittent switch operation

Firm Action

On 12/19/06 the firm sent letters via return receipt to all domestic customers. The letter advised customers of the issue and if the user encounters the problem, to press the button again with additional force. The users are to contact Philips for replacement. The devices can be used until the user receives a replacement.

Root Cause

Process design

Distribution Pattern

Distribution is worldwide.