HeartStart FR2+ Defibrillator
FDA Recall Z-0308-2007 · Posted 2006-12-27
Advertisement
Recalling FirmPhilips Medical Systems
Recall StatusTerminated
Product CodeMKJ
Date Initiated2006-12-19
Date Terminated2011-06-22
Product Quantity2000 units
Reason for Recall
Potential for shock treatment to be delayed or prevented due to board contamination that causes intermittent switch operation
Firm Action
On 12/19/06 the firm sent letters via return receipt to all domestic customers. The letter advised customers of the issue and if the user encounters the problem, to press the button again with additional force. The users are to contact Philips for replacement. The devices can be used until the user receives a replacement.
Root Cause
Process design
Distribution Pattern
Distribution is worldwide.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.