Arrow International StimuCath(R) Continuous Nerve Block Kit, ASK-19608-SFH, Lot # MF0127039 Permits placement of catheters next to nerves and nerve plexuses for continuous nerve block anesthesia of analgesia techniques for periods not exceeding 72 hours
FDA Recall Z-0309-2012 · Posted 2011-11-30
Advertisement
Recalling FirmArrow International Inc
Recall StatusTerminated
Product CodeCAZ
Date Initiated2011-06-20
Date Terminated2012-10-01
Product Quantity17 kits
Reason for Recall
Some of the outer trays used in the packaging of the ASK-19608-SFH Continuous Nerve Block Kits, lot # MF0127039 were found to have an insufficient seal. Therefore, the packaging may have been compromised, and sterility of the product cannot be guaranteed.
Firm Action
The firm, Arrow International, sent an "URGENT MEDICAL DEVICE RECALL" letter dated June 20, 2011, to its customer. The letter included a stock response form to be completed and returned to the firm. The letter also described the product, problem and action to be taken. The customer was instructed to check their stock for the products included in the scope of the recall; cease use and distribution; quarantine all affected product immediately; complete and return the enclosed Recall Acknowledgement & Stock Status Form to their sales representative, and return any affected product freight collect to: Arrow International, Inc., 312 Commerce Place, Asheboro, NC 27203, attn: Asheboro Quality Assurance Manager. Arrow International can provide replacement with a similar product or credit their account.
If you have questions or concerns, please call this toll free number 800-233-3187.
Root Cause
Packaging
Distribution Pattern
Nationwide distribution: CT