Terminated

Access 2 Immunoassay System, Remanufactured, Catalog Number 386220. In vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids.

FDA Recall Z-0309-2018 · Posted None

Recalling FirmBeckman Coulter Inc.
Recall StatusTerminated
Product CodeJJE
Date Initiated2017-11-06
Date Terminated2019-11-04
Product Quantity3 units

Reason for Recall

Potential for minor skin burn if the tool contacts the reagent storage interface board (B79917) when performing a pipettor alignment.

Firm Action

An Urgent Medical Device Recall letter, dated November 6, 2017, were sent to the affected customers via mail. The letter identifies the affected device, the reason for recall, impact, and actions to be taken. The firm states that customers may continue to operate their system normally. Customers are instructed to contact Technical Support if a device on the Access 2 instrument needs alignment. The alignment tool is only to be used by the Beckman Coulter service personnel and Beckman Coulter trained biomedical engineers. A Beckman Coulter service rep will contact customers to schedule a service visit. Customers are to share this information with their staff and retain a copy of the letter. Customers are to respond via one of the ways identified. Beckman Coulter should be contacted with any questions regarding the notice.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution -- US, Argentina, Canada, Chile, China, Columbia, Costa Rica, France, Georgia, Germany, Ghana, Greece, Hong Kong, India, Italy, Japan, Kazakhstan, Mexico, Morocco, Myanmar, Nepal, Philippines, Poland, Russia, Rwanda, Saudi Arabia, Slovakia, South Africa, South Korea, Thailand, Turkey, United Kingdom, and Vietnam.