Brand Name: Aesculap Product Name: SEALING UNIT F/10/12MM TROCARS W.REDUCER Model/Catalog Number: EK083P Software Version: N/A Product Description: SEALING UNIT F/10/12MM TROCARS W.REDUCER Component: N/A
FDA Recall Z-0309-2026 · Posted 2025-10-24
Advertisement
Recalling FirmAesculap Inc
Recall StatusOpen, Classified
Product CodeGCJ
Date Initiated2025-09-24
Product Quantity26 units
Reason for Recall
It was determined there is the potential of silicone fragments detaching from the yellow sealing component.
Firm Action
On September 24, 2025 URGENT MEDICAL DEVICE RECALL NOTIFICATION letters were sent to customers. Actions to be taken. 1. Review the recall notice and ensure all users in the organization are notified. Forward this notification to customers. Remove affected units from inventory, quarantine the product but do not destroy it. Complete and return the included response form. Return affected units to the firm. Credit of the product will be issued. An alternative disposable universal seal is available and can be ordered.
Root Cause
Under Investigation by firm
Distribution Pattern
Worldwide - US Nationwide distribution in the states of AR, AZ, CA, CT, MD, NE, NM, NY, OR, PA, WA, WI and the country of Canada.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement