Total Thyroxine (T4) ELISA; REF EIA-1781

FDA Recall Z-0310-2022 · Posted None

Recall StatusOpen, Classified
Product CodeKLI
Date Initiated2021-10-18
Product Quantity314 total (270 US)

Reason for Recall

Firm has identified two lots of product that do not meet product shelf life, resulting in failed Controls. The clinical impact may be a delay in diagnosis and a consumption of patient samples.

Firm Action

On 10/18/2021, DRG issued Medical Device Recall Notification letters and response forms to the End-user through confirmation of the Distributor. All customers were notified via e-mail. Actions Required: 1) Please immediately discontinue use or distribution of the affected lots. 2) If you have further distributed the affected product lots, please identify and immediately notify those customers at once of this product recall by providing a copy of the recall notice and response form. 3) If you have relabeled or rebranded the affected product lots, please follow your internal recall process and notify the appropriate regulatory bodies as required. 4) Disposal of all affected lots should be in accordance with applicable national, st ate, and local laws and regulations. 5) Complete and return the response form.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Domestic: NY; International: TZ, VN, BD, CO, TT, YE, BO, DE, SV, GE, PY, TG, RU/UA, NI, DO, NG, EG & CM