Terumo Sarns Level Sensor II Pads. For use with Sarns Advanced Perfusion System I, Perfusion System 800 and Perfusion System 900. Terumo Cardiovascular Systems Corp., Ann Arbor, MI.; REF 195240. Level Sensor pads are used to attach the level sensor to the sides of a rigid reservoir.

FDA Recall Z-0311-2009 · Posted 2008-12-10

Recall StatusTerminated
Product CodeDTQ
Date Initiated2008-01-13
Date Terminated2009-07-07
Product Quantity31,980 pads (25,620 to U.S. and 6,360 international)

Reason for Recall

The level sensor holder may detach from the adhesive pad causing the sensor to lose contact with the venous reservoir, which may result in a detached sensor indication or false alarm.

Firm Action

Consignees were notified via recall letter (Urgent Medical Device Removal letter) dated 2/13/08 and requested to report stocks on hand so that replacements could be sent.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution including USA and countries of Australia, Belgium, Canada, China, Dominican Republic, Japan, Saudi Arabia, Singapore and Yemen.