Terminated

BD Max Instrument Reader Spare Part, Catalog Number 8090369/443364, 44336409 (Repaired Reader spare part) - Product Usage: The BD MAX System, used in conjunction with appropriate extraction kits and reagents, is capable of automated extraction and purification of nucleic acids from multiple specimen types. The BD MAX System is also capable of automated amplification and detection of target nucleic acid sequences for use in clinical, industrial, and research settings.

FDA Recall Z-0312-2019 · Posted None

Recall StatusTerminated
Product CodePQA
Date Initiated2018-09-24
Date Terminated2020-04-10
Product Quantity23

Reason for Recall

A trend for false positive Candida glabrata results has been confirmed. The discrepant results impact the Candida glabrata target only, when in the presence of a Trichomonas vaginalis true positive.

Firm Action

Urgent Medical Device Product Recall notification letters dated 9/24/18 were sent to customers.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide distribution - US nationwide in the states of AL, AZ, CA, CO, DC, FL, GA, IL, IN, KY, LA, MD, MI, MO, NC, NJ, NM, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, WI, and Puerto Rico. Countries of Spain, Germany, France, Austria, Denmark, Switzerland, Sweden, Italy, Kuwait, Oman, Saudi Arabia, and Turkey.

Official Source

View FDA Recall →